N NEXUS MOLECULAR LABS ANALYTICAL SERVICES ISO/IEC 17025 Accredited · CAP Certified · Health Canada Licensed

Certificate of Analysis

DSIP (5mg)

Report No: NML-DSIP-260604-001 Analysis Date: 2026-06-21 Report Issued: 2026-06-25
Sample Identification
Compound DSIP (5mg) Client Numira Peptides  ·  numira.cc
CAS Number62568-57-4Date Received2026-06-17
Molecular FormulaC35H48N10O15PDate Tested2026-06-21
Molecular Weight 848.94 g/mol PubChem CID16129617
Lot Number NML-DSIP-260604 Test Methods HPLC-UV · ICP-MS · Kinetic LAL
Appearance White lyophilised powder Date Reported 2026-06-25
HPLC-UV Purity & Content Analysis
5mg
Ordered Amount
4.93mg
Measured Amount
98.72%
Purity (HPLC-UV)
98.6%
Recovery
Fig. 1 — HPLC-UV Chromatogram (214 nm)  ·  Run: NML_DSIP_DSIP-260604_1.d
mAU 0 300 600 900 1200 0 2 4 6 8 10 12 Time [min] DSIP 4.02 min
DSIP (5mg) vial
APPEARANCE
#Retention Time (min)Area% AreaAssignment
11.781.2E+020.61Unknown impurity
24.024.010E+0498.72DSIP
36.151.1E+020.47Unknown impurity
Bacterial Endotoxin Test (USP <85>)
MethodResultSpecificationVerdict
Kinetic LAL (USP <85>)< 0.125 EU/mL≤ 0.125 EU/mLPASS
Elemental Impurities (USP <233> / ICH Q3D)
ElementMethodResult (ppm)Limit (ppm)Verdict
Arsenic (As)ICP-MS< 0.0012.1PASS
Cadmium (Cd)ICP-MS< 0.0010.5PASS
Lead (Pb)ICP-MS< 0.0010.5PASS
Mercury (Hg)ICP-MS< 0.0013.0PASS

Note: Peptide samples may contain counter-ions (TFA, acetate) and excipients (mannitol, sucrose) used as lyoprotectants and pH buffers. These are not detected by UV and are not classified as impurities under USP <233>.

Analysis Performed by
Dr. M. Chen, Ph.D. (Analytical Chemistry)
Senior Analytical Chemist · Nexus Molecular Labs
info@nexusmolecularlabs.com
2026-06-25